Description
Sonotech Pro
Sonotech Pro doppler
Colour Display : Diagnostic Representations
FHR Wave and 3 modes
FHR Wave and 3 modes
Exchangeable probe 2Mhz, 3Mhz, 4Mhz, 8Mhz
Features
Standalone 2MHz Water Resistant (IPX4) Probe
FHR Display: Bar Graph | Histogram | Waveform
Working Diagnostics: Realtime, Average, & Manual
Out of Range Fetal Heart Rate Alarm Display
Backlit LCD Battery Status Indicator Display
+8 Hrs Operation on 2 AA Alkaline Batteries
Built In Speaker (=1.8W) with Volume Control
Audio Out Provisions: Headphones, PC Recording
Auto Power Off on No Signal (8 seconds) Function
FHR Measuring Diagnostic Range: 50~240BPM
Diagnostics Accuracy: } +/-2 Beats per Minute
Working Ultrasonic Frequency: 2.0MHz} 10%
Ultrasonic Opt Intensity: Isata < 5 mW/cm2
Ultrasonic Working Mode: Cont Wave Doppler
Available in two Colours- Blue and Pink
Ultrasonic pocket probes types:
2.0 MHz waterproof Ultrasonic probe
3.0 MHz waterproof Ultrasonic probe
4.0 MHz waterproof Ultrasonic probe
8.0 MHz waterproof Ultrasonic probe
Specifications
-
- Display: 45mm×25mm LCD display
FHR Performance
-
- FHR Measuring Range: 50~240BPM (BPM: beat per minute)
-
- Resolution: 1BPM
-
- Accuracy: ±2BPM
-
- Power consumption:<0.8W
-
- Auto Shut-OFF: After 1 minute no signal, power off automatically.
-
- Battery Type Recommended: Two pieces of 1.5 V DC battery (SIZE AA LR6).
Probe:
-
- Nominal Frequency: 2.0MHz, 3Mhz, 4Mhz, 8Mhz
- P-: <0.5MPa
- Working Frequency: 2.0MHz±10%
: <10 mW/cm
: <50mW/cm
-
- Ultrasonic Output Intensity: Isata<5 mW/cm
Ultrasonic Output Power: P<10 mW
Working Mode: Continuous wave Doppler
-
- Effective Radiating Area of Transducer: 157mm
-
- ±15%
-
- Temperature: +5℃~+40℃
-
- Humidity: ≤80%
-
- Atmospheric Pressure: 70 kPa~l06kPa